Best Practices for Pressure Instrument Calibration
|technical

Best Practices for Pressure Instrument Calibration

แรงดันสอบเทียบแนวทางปฏิบัติ
# Best Practices for Pressure Instrument Calibration Proper pressure calibration is essential for safety and quality in manufacturing processes. ## Pre-Calibration Steps 1. **Visual inspection** - Check for leaks, cracks, and fitting condition 2. **Temperature conditioning** - Allow instruments to acclimatize for at least 2 hours 3. **Purge air** - Remove all trapped air from the system ## Calibration Procedure 1. Perform at least 3 pre-exercise cycles to reduce hysteresis 2. Calibrate both ascending and descending, minimum 5 points 3. Record all readings carefully 4. Calculate measurement uncertainty ## Precautions - Never exceed the maximum range of the unit under test - Use appropriate media: air for low pressure, oil for high pressure - Verify traceability of reference standards